AUTOLOG™ - Equipos y descartables para autotransfusión

Sophisticated, Effective Processing System

• The autoLog™ Autotransfusion System´s fully automated design design consistently produces a high-quality end product.

• The system’s preset operating program completes one processing cycle (fill, wash, empty) in approximately three minutes and returns approximately 135 ml of blood to the holding bag.ç

• The autoLog® System uses a two-strange filling process capable of first stage fill speed of 600 ml/min followed by a second stage fill spedd at either 600 or 250 ml/min based on incoming volume.

• While a processing cycle can be initiated manually, the autoLog® System has a unique self-start feature


A Source of High-Quality Blood Product

• The autoLog™ System is designed to consistently produce a blood producto with a hematocrit of 50% or greater.

• The autLog™ System’s unique bowl design packs blood tightly to produce an increased hematocrit and allows blood to enter at a moderate G-force to minimize hemolysis. The end product remains consistent and is independent of the type of surgical procedure.


• The variable speed wash process, coupled with the unique design of the bowl, promotes the effective removal of free-plasma hemoglobin, residual anticoagulant agents, activated platelets, white blood cells, and activated clotting factors of 90% of grater in accordance with AABB guidelines.
Simplicity of Operation

• Continuous software monitoring improves quality of end product and speed of processing.

• Easy system set-up and intuitive feature design increase efficiency and save time for system operators.

• The autoLog™ System is an all-inclusive processing system, saving you money,storage space and time on inventory management.

Why Use Autologous Red Blod Cell Salvage?

Autologous red blood cell salvage is a safe, reliable, cost effective method of returning red blood cells back to patients. The autologous red cell salvage procedure maximizes patient safety, overall cost savings, and blood quality. It:

• Eliminates the risk of clerical errors during blood handling

• Minimizes costs associated with the use of allogenic blood transfusion.

• Minimizes the use of banked blood products, thereby reducing the risk for transmission of blood-borne diseases, lowering the risk of transfusion reaction, and providing and option for those with religious objections to transfusions.

• Addresses the issue of blood shortages. Blood shortages are common and will become more acute in the next few years.

• Provides red blood cells with normal 2,3 DPG levels and viability superior to preoperative autologous blood donations.

Facts on Blood Prices and Supply

• Blood product prices have increased significantly due to the rising costs of safety measures and requirements for the blood supply.

• National blood inventory levels have dropped well below a safe and adequate supply. Certain critical blood types are not routinely available and as a result some elective surgeries have been postponed or cancelled.

• Transfusion for cardiac surgery utilizes approximately 20% of the United States’ blood supply.

STSTEN SPECIFICATION

Electrical Classification
Class I, Type B, Ordinary, Continuous Operation

Power

Voltage
110/120 of 220/240 V
Cycles
50-60 Hz
Phase
Single
Current
1.6/0.8 amps (depending upon voltage selection)
Fuses
4 AT/240 V
Power cord

2 wires plus ground (earth) connector;
3-prong hospital grade (USA only)

Speed and Rate Specifications

Centrifuge
0-10,000 ROM (+- 5%)
Pump
0-600 ml/min (+-5%)
Vacuum
200-280 mbar

Dimensions (Weight 32 kg (70 lb.))

Width
33 cm (13 inches)
Heighy
75 cm (30 inches)
Depth
22 cm (9 inches)

Temperatura Limit

Operacional
10ºC-30ºC (50ºF-86ºF)
Stotage
5ºC-30ºC (41ºF-86ºf)

HumidityRange

Operational
10-95% non-condensing
Stotage
10-95% non-condensing
DEMEDIC S.A. • Potosí 4012 (C1199ACH) • Ciudad de Buenos Aires • Argentina • (54-11) 4958-0536 • 0800-333-DEME (3363 ) • info@demedic.com.ar
Cerrar Ventana Subir